Warning: The magic method Gallery_Video::__wakeup() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/gallery-video/video-gallery.php on line 86

Warning: The magic method Hugeit_Slider::__sleep() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/slider-image/slider.php on line 262

Warning: The magic method Hugeit_Slider::__wakeup() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/slider-image/slider.php on line 264
Valcom VW2 Series | TKK Corporation


Valcom VW2 Series

  • Multiple process inputs: 4~20mA, 0~5V, 1~5V and 0~10V
  • Power supply of sensor is built-in.(DC24V Capacity: 50mA)
  • Models with analog outputs are multiple output type: 0~10V, -10~10V, 1~5V and 4~20mA.
  • Full high-intensity LEDs.
  • CE-certified.
  • Alarm setting 2 points(Three relay output or Photo-coupler Output)
  • Pattern select function included (Up to eight patterns)
  • Peak hold functions
    “sample hold”, “Peak hold”, “Bottom hold” and “peak-to-peak” functions included.
  • Various functions
    “Leading zero suppress”, “Forced-zero” and “Reelay reset” functions included.

Specification

Input Method Single-end
A/D Conversion Method ΔΣ conversion method
Sampling Rate MAX. 250 times / sec.
Display Main display:
Red/Green   7 segments LED display(Character height: Approx. 14.9mm)
Sub display:
White   7 segments LED display(Character height: Approx. 9mm)
Polarity Display Displayed automatically for negative results.
Display Range -19999~99999
Zero Display Leading zero suppress
Outer Control 4 out of the following 5 functions are assigned to the external control terminals.
* If the pattern select function is included, the remaining terminals available for other functions will be limited.
2 more functions (if 4 patterns used) or 1 more function (if 8 patterns used)
①Pattern select
②Hold
③Peak hold / Bottom hold / Peak to peak (Difference between the peak and the bottom)
④Forced-zero
⑤Relay reset
Range of Use Temperature -5~50°C
Range of Use Humidity 35~85%RH(No condensation)
Storage Temperature Range -10~70°C
Storage Humidity Range 60%RH or less
Power AC100~240V±10% 50/60Hz
Power Consumption AC100V: 12VA max.
AC240V: 15VA max.
External Size 96mm(W)×48mm(H)×99.7mm(D)
Mass Approx. 250g
Dielectric Strength Between Power terminal and Input terminal / External control terminals /Analog output   AC2000V for 1 minute
Between Power terminal and Comparative output   AC1500V for 1 minute
Between Input terminal and External control terminals/ Analog output / Comparative output   AC1500V for 1 minute
Between Case and Terminal   AC2000V for 1 minute
Insulation Resistance 100MΩ min. at 500V DC between above terminals
IP Rating IP66(Front)
Case (Material / Color) Polycarbonate / Black UL94V-0
Overrange Display OVER or –OVER for any input signal outside the display range.
“OVER” or “-OVER” is displayed if input exceeds the measurement range by more than ±10%.
Decimal Point Can be set at any desired place.

Process signal input specifications

Measuring Range 1~5V, ±5V, ±10V, 4~20mA, ±20mA
Display Offset: -19999~99999   Full scale: -19999~99999 Resolution: ±19999
Input Impedance Approx. 1MΩ: (1~5V, ±5V, ±10V)
Approx. 10Ω: (4~20mA, ±20mA)
Maximum Safe Input ±100V: (1~5V, ±5V, ±10V)
±50mA: (4~20mA, ±20mA)
Accuracy * ±(0.1% of FS +1digit) (At the time of 23℃±5°C and 35~85%RH)

* Accuracy applies when the sampling rate is 60 samples/s or less.

Comparison output specifications

Control Method Microcomputer-based
Trigger Value Range -19999~99999
Hysteresis Can be set in the range of 1 to 9999 digits for each value.
Comparison Operation Depends on the sampling rate.
Setting Condition H. G. L.
AL1 judgment value>AL2 judgment value(GO)>AL3 judgment value
Comparison Condition Judgment
Displayed value>AL1 judgment value AL1: ON
AL1 judgment value≧Displayed value≧AL3 judgment value AL2: ON
AL3 judgment value>Displayed value AL3: ON
G. L. L.
AL1 judgment value(GO)>AL2 judgment value>AL3 judgment value
Comparison Condition Judgment
Displayed value≧AL2 judgment value AL1: ON
AL2 judgment value>Displayed value≧AL3 judgment value AL2: ON
AL3 judgment value>Displayed value AL3: ON
H. H. G.
AL1 judgment value>AL2 judgment value>AL3 judgment value(GO)
Comparison Condition Judgment
Displayed value>AL1 judgment value AL1, AL2: ON
AL1 judgment value≧Displayed value>AL2 judgment value AL2: ON
AL2 judgment value≧Displayed value AL3: ON
Comparison Relay Contact Rating AC125V 0.3A (Resistive load)
DC30V 1A (Resistive load)
Number of Contacts Relay contact×3 (VW27N-2S□)
Minimum Applicable Load 10µA   10mV   DC
Mechanical Lifespan More than 50 million times
Electrical Lifespan More than hundred thousand times (Resistive load)
Photo-coupler Output
Open Collector Output (NPN)
Output Constant Sink current: 50mA MAX.
Applied voltage 30V MAX.
utput saturation voltage 50mA: 1.2V or less
Number of Outputs Photo-coupler Output (NPN) ×3 (VW27N-2SO□)
Comparison Warning Function Normal judgment output, zone judgment output, tolerance judgment output
Comparison Condition Memory 8 patterns are memorized.

Analog output specifications

Load Resistance 0~2V, 0~10V, -10~10V, 1~5V:10kΩ or more
0~20mA, 4~20mA:550Ω or less
Accuracy ±(0.1% of FS)
Ripple 0~2V, 0~10V, -10~10V, 1~5V:±50mVp-p
0~20mA, 4~20mA:±25mVp-p
Conversion Method D/A conversion method
Resolution Equivalent to 15 bit (Depends on the scaling setting)
Scaling Digital scaling
Response Speed 10ms or less   0→90% Responsed (Sampling rate: 250 times/ sec.)
* Response speed: 2ms±2 (1 / Sampling rate )ms or less