Valcom VSW2 Series (For Minute-Low-Absolute pressure) | TKK Corporation


Valcom VSW2 series

(For Minute-Low-Absolute pressure)

  • “Max./min memory” function included, which allows abnormal pressures detected so far to be checked during operation.
  • Color LCD adopted with display colors “red and green” intuitively understandable for users.
  • Compact design (length: 111 (107) mm, width: 37 mm) suitable as a replacement for mechanical switch.
  • Setting output can be selected between higher/lower limits and NPN/PNP (Factory-set at NPN/PNP.)
Measured medium M sensor: Dedicated to inert gases (noncorrosive gases) such as air, nitrogen, argon.
H sensor: Water, oil, gases or other media that will not corrode Hastelloy C22 and SUS316L.
Pressure type Positive, absolute, and compound pressures can be measured.
Pressure range M sensor: 5kPa
H sensor: 50~700kPa and 50~700kPa abs

* EAC Mark is an option item: EAC Mark option includes TECHNICAL RULES OF THE CUSTOMS UNION (TR-CU) Declaration-of-conformity document only, or includes both of TECHNICAL RULES OF THE CUSTOMS UNION (TR-CU) Declaration-of-conformity document and Metrology Pattern Approval certificate (MPA).
* The products in the photo are standard specification models. Please contact our sales office for further details.

Specification

Display -1999~2000(*1) 7 segments 4 digit LCD(Character height:9mm)
Display Color Red or Green (Arbitrary setting)
Rated Capacity 5, 50, 100, 200, 350, 500, 700kPa
50, 100, 200, 500, 700kPa abs
Calibration Range 0~(5, 50, 100, 200, 350, 500, 700)kPa
0~(50, 100, 200, 500, 700)kPa abs
Accuracy ±0.5%F.S.±1digit
(Non Linearity, hysteresis and repeatability included.)
Output Channels 2 points of Open collector output
1 points of Open collector output + 4~20mA output
Maximum rating for setting output: DC30V 20mA Residual voltage:0.85V(NPN output)
+Analog output Load resistance: 400Ω or less
Use Temperature Range Temperature: 0~50°C(Temperature of medium to be measured: 0~70°C)
Safe Overload Rating 300%R.C.(Semiconductor thin-film type sensor)
200%R.C.(Semiconductor sensor)(*2)
IP Rating Equivalent to IP65(Dedicated cable connection state)
Temperature Effect ±2.5%F.S.(25°C standard, within the service temperature range)
Vibration-Proof 98m/s²(10G)[Compliant with IEC68-2-6]
Connection Screw R1/4[Male screw]
RoHS Europe RoHS directive conformity product
Connector Model number XS2M-D423 made in Omron
Measurement Medium M sensor: Dedicated to inert gases (noncorrosive gases) such as air, nitrogen, argon.
H sensor: Water, oil, gases or other media that will not corrode Hastelloy C22 and SUS316L.
Power Source Voltage DC18~30V
Consumption current Approx. 30mA(Approx. 50mA for models with 4~20 mA output)
Functions Peak hold / Damping /FIX zero / Maximum and Minimum memory
Window Comparator / Zero shift / Key protection
Material Connection screw: Aluminum 2017(Pressure port/Semiconductor)M sensor
Connection screw: SUS316LH sensor
Diaphragm: Hastelloy C22H sensor
Weight M sensor: Main body Approx. 160g Dedicated cable 2m Approx. 60g
H sensor: Main body Approx. 200g Dedicated cable 2m Approx. 60g
Accessories Operation manual (1 copy)(Specify English or Japanese.)
Test report (1 copy)
Options (*3) Wall hangings fixation bracket [VSW2-BR]
  • (*1) Depends on the factory setting. Values beyond the calibration range cannot be displayed.
  • (*2) R.C.: Rated Capacity
  • (*3) Contact our sales office for details.