Valcom VCCP Series | TKK Corporation


Valcom VCCP Series

  • Measurement of the high vacuum area is possible.
    0~1500Pa(abs), 0~5kPa(abs) and 0~10kPa(abs)
  • High leak performance
  • Maximum allowable pressure (400kPa)
Measured medium Water, oil, gases or other media that will not corrode Ceramic + SUS316L + O-ring.
Pressure type Compound pressure can be measured.
Measured Range Can be specified within the range of 1500Pa(abs), 5kPa(abs) and 10kPa(abs) of calibration.
Output type Specify among 2 to 20mA(4 wire system), 1 to 5V

Specification

Non Linearity ±0.8%F.S. or less
Hysteresis ±0.8%F.S. or less
Range of Use Temperature 0~70°C (No condensation or freezing)
Temperature Effect on Load ±0.5%F.S. / ±10°C or less
Safe Overload Rating ±0.5%F.S. / ±10°C or less
Connection Screw Material SUS316L
Sensor Element Alumina ceramic
O-ring Material Selected from EPDM or Fluororubber (Class 4D) or Perfluoro-Elastomer
Case Material SUS304
IP Rating Equivalent to IP54
Power DC24V (±10%)
Mass Approx. 470g
Accessories Operation manual (1 copy), Test report (1 copy)