RADAR TYPE LEVEL INDICATOR | TKK Corporation


Feature

RADAR TYPE LEVEL INDICATOR(Frequency Modulated Continuous Wave)
VG7 Series

The radar type level meter can be used for various applications since the microwave speed is not affected by temperature, pressure, change in the permittivity of the measured substance, or density modulation.

  • It can be used for powders, granules, liquids, slurries and viscous substances.
  • Change of the measured substance does not affect the measurement.
  • Maximum of 40 m can be measured without contact.
  • It can be used under vacuum, high temperature and high pressure.
  • ATEX EExia (Intrinsic Safety Explosion-proof) certified products
  • Low power consumption (75 mW)

Operating Principles

RADAR TYPE

The frequency when the transmitted electromagnetic wave reflects and reaches the receiver is detected. The required time is calculated from the change portion, converted into the level position, and then output.
For frequency modulation, signal processing to keep linearity is programmed.

Standard Specification

Power Supply
(Two Wire)
Non Explosion-proof・EExi DC24V(DC14~30V)
EExd DC24V(DC20~36V)
Measuring Range Min Tank Height 0.5m
Max Measuring Range 40m(dependant on environment)
Blocking Distance Extended Length + Antenna Length + 0.1m
Output Output 1
(Max Load Resistance 350Ω)
DC4~20mA HART
(Max Load Resistance 350Ω)
Error Output DC3.6mA or DC22mA or hold
Accuracy 0.05%
Resolution ± 2μA
Accuracy Accuracy ±3mm(dependant on environment)
Resolution 1mm
Beam Angle(DN50) 15°
Beam Angle(DN80) 10°
Conditions Ambient Temperature -40~+80℃(EExia:-40~+60℃)
Storage Temp. for Insutrment -40~85℃ 
Flange Temperature -40~+200℃(EEx:+150℃)
Withstanding Pressure 0.1MPa~4MPa
Enclosure Rating IP-66 NEMA6-6X

Product Lineup


VG7

VG7 DN50 Standard Type


VG7

VG7 DN50 Extension Type


VG7

VG7 DN80 Standard Type


VG7

VG7 DN80 Extension Type

In addition to the above, various specifications are available.