Warning: The magic method Gallery_Video::__wakeup() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/gallery-video/video-gallery.php on line 86

Warning: The magic method Hugeit_Slider::__sleep() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/slider-image/slider.php on line 262

Warning: The magic method Hugeit_Slider::__wakeup() must have public visibility in /home3/cp504663/public_html/product/wp-content/plugins/slider-image/slider.php on line 264
Valcom P3MB Series | TKK Corporation


Valcom P3MB Series

Make to order

Excessive High Pressure Transducer (High Accurate)

  • Ultrahigh pressure: 3000bar / High accurate: 0.2%R.C.
  • Good reliability and stability.
  • Measurement of changes in static/dynamic pressures, peak pressure and pressure fluctuation.
  • Protection rating IP67 (Compliant with DIN 40050, IEC 529)
Measured medium Water, oil, gases or other media that will not corrode SUS630.
Pressure type Positive pressure can be measured.
Rated capacity 3000bar

Specification (Mechanical input characteristics)

Pressure type Absolute pressure
Initial Value 0bar
Mechanical Values to VDI/VDE2600 by Reference to Full Scale Value Operating Range at Reference Temperature 0~150%
Overload Limit at Reference Temperature 200%
Test Pressure 150%
Destructive Range >200%
Max. Dynamic Pressure 100%
Max. Dynamic Vibration Amplitude to DIN 50 100 70%
Natural Frequency of the Membrane >100kHz
Dead Volume 900mm³
Control Volume 1.5mm3
Material of the Parts In Contact with the Measurement Medium 1.4542
of the Parts In Contact with Environment 1.4542, 1.4301, Nickel plating brass
Chloroprene natural rubber, Silicon

*This is a made-to-order model. Contact our sales office for details.

Specification (Output characteristics)

Signal Span(Sensitivity) 1.5±0.15%mV/V
Characteristic Curve Deviation ±0.2%
Repeatability to DIN1319 ±0.05%
Temperature Effect on Load (*1) In the Norminal Temperature Range ±0.1%
In the Operating Temperature Range ±0.2%
Temperature Effect on Zero Balance (*1) In the Norminal Temperature Range ±0.1%
In the Operating Temperature Range ±0.15%
Normal(Rated)Range 0.5~12V
Input Resistance at Reference Temperature 350±5Ω
Output Resistance at Reference Temperature 350±1.5Ω
Isolation Resistance at 500V AC 5000MΩ

(*1) Effect of temperature change per 10k within the rated range of bridge applied voltage (Percentage of measured value)

Specification (Ambient environment)

Reference Temperature +23°C
Nominal Temperature Range -10~+80°C
Limiting Temperature Range -40~+100°C(Up to 120°C within 24 hours)
Storage Temperature Range -50~+100°C
IP Rating (DIN 40050, IEC 529) IP67

Specification (Mechanical specification)

Pressure Connection M20×1.5
Weight (Approximation) 200g(Cable excluded)
Accessories Double cone seal (×1)